|
|
||||||||
Soil and Water Management Research Unit, Conservation & Production Research Laboratory, USDA-ARS, Bushland, TX 79012
* Corresponding author (srevett{at}cprl.ars.usda.gov)
1 The mention of trade or manufacturer names is made for information only and does not imply an endorsement, recommendation, or exclusion by USDA-Agricultural Research Service. ![]()
Received 21 March 2005.
| ABSTRACT |
|---|
|
|
|---|
Abbreviations: BEC, bulk electrical conductivity EM, electromagnetic IAEA, International Atomic Energy Agency NIST, U.S. National Institute of Standards and Technology NMM, neutron moisture meter RMSD, root mean squared difference TDR, time domain reflectometry
| INTRODUCTION |
|---|
|
|
|---|
Conventional TDR systems are here defined as those that employ probes (usually bifilar or trifilar waveguides) that are inserted into or buried in the soil and are connected to a TDR instrument either directly or through a system of coaxial multiplexers, for the purpose of capturing a waveform that is analyzed to determine the travel time of the TDR pulse along that length of the probe rods that is in contact with the soil. Due to installation difficulties, these TDR systems are not well adapted for determination of water contents well below the root zone. However, a conventional TDR system was used in the soil column comparison study as a reference system because of its perceived relative immunity to temperature and bulk electrical conductivity interferences, with the intention of using water contents determined by TDR at multiple levels in each column to calibrate the other instruments. Calibration of the other systems depended on good accuracy with the TDR system. Here we discuss calibration of the conventional TDR system.
The TDR method makes use of electrical theory for signals in wave guides. For a coaxial cable, the value of the propagation velocity, v, of an electronic pulse along the cable is inversely proportional to the permittivity,
, of the dielectric (insulating medium, often plastic) between the inner and outer conductors of the cable:
![]() | [1] |
a, which in turn affects the velocity of a pulse along the probe rods. The measured property in the TDR method is the travel time, tt, of the electronic pulse along the length (L) of the probe rods that are exposed to the soil. The velocity of the pulse can be calculated as v = 2L/tt. Assuming µ = 1, one sees that an apparent permittivity,
a, may be determined for a probe of known length, L, by measuring tt:
![]() | [2] |
v, and values of
a determined from Eq. [2] for four mineral soils.
![]() | [3] |
Since 1980, other researchers have shown that the relationship between
v and tt/(2L) is practically linear (e.g., Ledieu et al., 1986; Yu et al., 1997). Indeed, Topp and Reynolds (1998) found that Eq. [3] is equivalent to
v = 0.115(
a)0.5 0.176. We note here that the apparent permittivity, as calculated from travel time using Eq. [2], contains any deviation from unity of µ. In addition, the value of
a increases with the bulk electrical conductivity,
a (S m1), of the soil (Wyseure et al., 1997; Robinson et al., 2003), particularly for
a >0.2 S m1. Also, the value of
a increases with soil water content (Rhoades et al., 1976; Mmolawa and Or, 2000). The value of
a may increase or decrease with temperature depending on the soil texture (Campbell, 1990; Pepin et al., 1995; Persson and Berndtsson, 1998; Wraith and Or, 1999) and increases as measurement frequency decreases (Campbell, 1990). The latter fact means that, for a broadband method such as TDR, there is a cable length effect because coaxial cable acts as a low pass filterthe longer the cable, the less signal energy is present in the higher frequencies. The TDR estimated value of
a increases with cable length (Hook and Livingston, 1995), particularly for high surface area soils (Logsdon, 2000). Topp et al. (2000) found that TDR signal dielectric loss is a function of
a, regardless of whether this conductivity arises from soil water solution conductivity or from clay type and content. Thus, TDR calibrations should take
a into account, and probably cable length as well.
Fortunately, conventional TDR may be used to assess
a (Wraith, 2002)
![]() | [4] |
o is the permittivity of free space (8.854 x 1012 F m1), co is the speed of light in a vacuum (299792 458 m s1), L is the probe length (m), Vo, VF, and VI are relative voltages measured from the wave form (Fig. 1)
, Zo is characteristic impedance of the probe (
), and Zu is the characteristic impedance of the cable tester (
). Topp et al. (2000) and others found that Eq. [4] accurately provides the soil BEC. Thus, it should be possible to include the important effects of temperature-dependent
a in a soil specific TDR calibration.
|
| HYPOTHESIS DEVELOPMENT |
|---|
|
|
|---|
, the effect of direct current electrical conductivity,
dc, on the apparent permittivity can be represented by (Robinson et al., 2003)
![]() | [5] |
' is the real component of the complex dielectric permittivity,
''relax is the increase in permittivity due to relaxation losses, and the other terms are as defined above. Although Eq. [5] is for a single frequency, it includes the effects that are important interferences to the TDR method. As the signal frequency decreases (longer cables), the value of
dc/
increases, leading to larger values of
a. As conductivity increases (soils with larger BEC), the value of
a increases, more so at lower frequencies. As relaxation losses increase (e.g., bound water effects), the value of
a increases. For broad band signals such as that of TDR, the angular frequency may be replaced by 2
f, where f is an effective frequency (Robinson et al., 2003), which previously has been calculated for TDR in at least two different ways (Or and Rasmussen, 1999; Topp et al., 2000).
Parallel with the increase of travel time with the square root of permittivity (Eq. [2]), Eq. [5] suggests that travel time will increase with the square root of conductivity. Assuming that relaxation losses are not large, we hypothesize that, for situations in which the effective frequency is unvarying, a useful calibration model will be
![]() | [6] |
The effective frequency will decrease for longer cables and in dispersive soils. Also, the degree of dispersion may increase with water content (Dirksen and Dasberg, 1993; Robinson et al., 2003), leading to a decrease in the effective frequency as water content increases. Not all clay soils are dispersive, as was illustrated by Evett (2000b), who contrasted TDR waveforms across a range of water contents for the nondispersive, kaolinitic Cecil clay vs. the dispersive Pullman soil. Equation [5] also suggests that travel time will vary as the square root of 1/f. We hypothesize that in dispersive and conductive soils, or in systems with varying cable lengths, a useful calibration model will be
![]() | [7] |
''relax, the increase in permittivity due to relaxation losses. The electromagnetic (EM) methods (time domain and capacitance or frequency domain) of soil moisture sensing typically allow data logging and unattended operation, but with uncertain precision and accuracy in soils of the U.S. southern Great Plains (Baumhardt et al., 2000; Evett and Steiner, 1995). U.S. agricultural soils in the Great Plains and further west often exhibit three important horizons differing in texture and/or chemical composition: (i) a well mixed Ap horizon, (ii) an illuvial clay horizon below that featuring larger clay content, and (iii) a horizon of carbonate accumulation below that, sometimes containing 50% or more CaCO3 along with some CaSO4. These contrasts in texture and chemical composition could potentially affect the calibrations of soil water content sensors. Our objective was to calibrate conventional TDR in three soil materials representing one instance of these important horizons, determining the accuracy, differences among soils, and sensitivity to temperature, bulk electrical conductivity, and cable length. Comparisons were made vs. soil water content determined by mass balance in soil columns. If accurately calibrated, TDR determined water contents would be used to cross-calibrate the other sensors in our larger study.
| MATERIALS AND METHODS |
|---|
|
|
|---|
Each soil was packed uniformly in 5-cm lifts into three replicate columns. Soil in each column was 75 cm deep and 55 cm in diameter, and rested on a 5-cm-deep drainage bed of fine pure silica sand in which a ceramic filter tube specified at 100-kPa air entry potential was embedded. Soil was packed around access tubes for other sensors, which were held in place with a jig so the tube positions would be identical in each column. For the conventional TDR systems, horizontal, trifilar TDR probes (model TR-100, 20-cm length, Dynamax, Inc., Houston, TX) were installed at depths of 2, 5, 15, 25, 35, 45, 55, and 65 cm in each column, and thermocouples were installed at the same depths to measure soil temperature. Three samples for initial gravimetric water content were obtained every two layers. Column sides were covered with reflective aluminum foil to minimize diel heating and cooling on the sides. Column soil surfaces were left exposed to solar radiation and air temperature variations in the greenhouse that housed the experiment. For measurements at the saturated end, the soil surface was temporarily covered with a sheet of polyethylene after excess water was suctioned from above the surface so the measured mass would not be increased by water standing on the surface.
Column mass was measured every 6 s using a data logger (model CR7, Campbell Scientific, Inc., Logan, UT) connected to the paralleled output of the four load cells in each deck scale (model DS3040-10K, Weigh-Tronix, Inc., Fairmount, MN), using a six-wire bridge configuration to minimize temperature-induced errors. Mean values were output every 5 min. Calibration with test masses traceable to the U.S. National Institute of Standards and Technology (NIST) resulted in RMSE values of linear regression
50 g for all scales. Initial volumetric water content of each column was computed from the mass of soil added, the volume of the column, and the water contained in the soil as determined from the gravimetric samples. The CR7 data logger was also used to acquire soil temperature data from the thermocouples. After an initial period during which data at the air-dry end were collected, the columns were infused with CO2 through the filters in the bottom of each column, followed by slow wetting through the filters under a hydraulic head of 2.2 m. It required several months to fully saturate all the columns. Because there was some soil swelling, the height of each soil column was measured periodically, and volume adjustments were made accordingly.
Measurements of travel time in the 72 20-cm trifilar TDR probes were made every 30 min using the TACQ program (Evett, 2000a, 2000b) running under DOS and controlling a conventional TDR system comprising an embedded computer (IBM PC/AT compatible), cable tester (model 1502C, Tektronix Inc., Redmond, OR), and five coaxial multiplexers arranged in a star configuration with one primary and the others secondary (Evett, 1998). Travel times were determined automatically by TACQ using the default waveform interpretation algorithms. Apparent dielectric constant was calculated using Eq. [2]. Total coaxial cable length varied among columns from 6.4 to 10.0 m, such that no one soil type had a preponderance of shorter or longer cables (Table 1). Bulk electrical conductivity was calculated from Eq. [4], using relative voltage values Vo and Vf determined using the waveform positions described in Evett (2000a)(2000c) (Fig. 1). For BEC calculations, the mean probe characteristic impedance for three probes was determined from repeated (n = 8) measurements of Vo and Vmin in deionized water using (Wraith, 2002)
![]() | [8] |
w is the permittivity of water, and Vo and Vmin are as in Fig. 1. Water temperature was measured using a thermometer traceable to NIST, and water permittivity was calculated according to Weast (1971)(p. E-61). Probe characteristic impedance measurements were repeated for each total cable length reported in Table 1, and with multiplexers included in the circuit.
|
Vt, of the second rising limb of the waveform, which represents the reflection of the TDR pulse at the end of the probe rods, and from the magnitude of the initial voltage step (TDR pulse height). The TACQ program was modified to output the slope value and the time base, tb (ns per unit), of the waveform (version TACQbeta available at http://www.cprl.ars.usda.gov/programs/, verified 18 Aug. 2005), and the magnitude of the initial voltage step was calculated as V0 VI from the BEC data output by TACQ (Evett, 2000a, 2000c) (Fig. 1). This differs from the procedure used by Topp et al. (2000), which relied on finding the maximum value of the second rising limb to fit a horizontal line tangent to it. Finding this maximum value may be difficult due to multiple reflections in the waveform. Also, this maximum value decreases as BEC increases, leading to a reduction in the reflected pulse magnitude. The resulting reduced rise time causes the effective frequency determined by the method to Topp et al. (2000) to be larger than that determined by our method, in effect confounding the effect of BEC on frequency (slope of the reflection) with the effect that BEC has on the magnitude of the reflected pulse (a conduction effect).
The effective frequency (radians) used in the present study, with subscript "vi" to indicate that it was based on the initial voltage step, was
![]() | [9] |
Assuming that calibrations of TDR travel time vs. water content are practically linear, an accurate two-point calibration should be possible if conductivity and temperature effects are minimal. Thus, the TDR system was calibrated vs. the column mean water contents for each soil using data from the air-dry state and the saturated state. In addition, calibrations were conducted with both travel time and conductivity as independent variables as in Eq. [6], and with travel time, conductivity, and effective frequency as independent variables as in Eq. [7].
| RESULTS AND DISCUSSION |
|---|
|
|
|---|
|
v vs. cott/(2L), where tt and L are measured. Combining data for the three soils, TDR calibration was highly significant (Table 3), with slope and intercept close to the values reported by Topp and Reynolds (1998), but with a slightly smaller slope value than those reported by Yu et al. (1997). The RMSE of regression value of 0.02 m3 m3 is comparable to the ±0.02 m3 m3 accuracy typically claimed for TDR. Moreover, multiple comparisons of intercepts and slopes (SAS, 2004) showed that these were significantly different for the three soils. Accuracy for the individual soil calibrations was better than 0.01 m3 m3 as indicated by the RMSE of regression. For the clay-rich A and B soils, calibration equation slopes were smaller than that reported by Topp and Reynolds (1998), which was based on the four mineral soils studied by Topp et al. (1980), three of which had clay contents in the range of the A and B soils studied here. The difference is probably attributable to differences in clay mineralogy, the clay in our soils being rich in smectitic clay (montmorillonite), which is known to be electrically more lossy than clays with smaller ion exchange capacities and surface areas. The smaller slopes for our A and B soils would mean that a given measured travel time would result in an overestimate of water content (for large water contents) if the Topp and Reynolds (1998) equation were used. However, errors that would occur for measurements in our Soils A and B using the equation of Topp et al. (1980) rather than our calibration equations are within ±0.02 m3 m3 for the range of water contents studied (
0.050.49 m3 m3). For our Soil C, use of the Topp et al. (1980) equation would underestimate water content by 0.042 m3 m3 at saturation, but by only 0.001 m3 m3 at air-dry water contents. Our Soil C is approximately 50% CaCO3, far from the kind of mineral soils studied by Topp et al. (1980).
|
|
|
|
|
dc/
becomes larger as
becomes smaller. Temperature dependency also increases with the BEC of the soil, which in turn increases with soil water content. This explains why, in the air-dry state when BEC was smallest and effective frequency was largest, none of our three soils exhibited any temperature dependency for TDR. Mindful that coaxial cables act as low pass filters and that increasing cable length will lead to loss of high frequency components of the TDR pulse, we regressed column mean water contents for each column vs. column mean temperatures and plotted the slopes vs. total cable length for each column (Fig. 5) . There is an apparent nonlinearly increasing temperature dependency with cable length. Thus, for Soils A and B it appears that a complete TDR calibration should account for the effects of both soil temperature and cable length. Since the effect of soil temperature is tied to the temperature dependency of BEC, and since the cable length effect is tied to the fact that cables act as low pass filters, then Eq. [6] and [7] may be useful calibration models.
|
for cable lengths ranging from 6.4 to 10.0 m, respectively (Fig. 6)
. For the same five cable lengths, impedance standard deviations ranged from 1.7 to 2.8
, in no particular order, indicating good repeatability among probes. In air-dry soil, relationships between
a and temperature were significant but weak (Table 5). Linear regression slopes were less than 1.15 x 104 and regressions explained less than 26% of the variation in
a. While the coefficients of determination were small, there was very little scatter in the data, as shown by the small values of RMSE. Mean values of
a were <0.042 dS m1, and
a increased with increasing clay content, in agreement with the results of Rhoades (1981), who found that soil matrix conductivity increased with clay content.
|
|
a were an order of magnitude larger, in agreement with the well known positive relationship between
a and water content (Gupta and Hanks, 1972; Rhoades et al., 1976; Bohn et al., 1982). Also,
a linearly increased with temperature for the saturated A and B soils, with linear regressions explaining 82% of the variation in
a (Table 5). The larger mean value of
a and larger slope for Soil B were expected due to this soil's greater clay content. The values found for Soils A and B are similar in magnitude to those found by Persson and Berndtsson (1998) for a mix of montmorillonite clay and sand and for a clayey moraine soil. Although the relationship between
a and temperature was apparently linear for Soil C (Fig. 7) , the coefficient of determination and slope were both smaller than for Soils A and B. The values of
a were also approximately twice as large for Soils A and B as for Soil C at any given temperature. These results explain both the results on the temperature dependency of travel times (apparent water contents) and the results of the calibrations. The latter showed lower calibration slopes (greater sensitivity of tt to changing water content) for Soils A and B, corresponding to the larger BEC values and greater temperature dependency of BEC for these soils. Note that the slopes for saturated soils in Table 5 are larger for Soils A and B than the well-established relationship between electrolytic conductivity and temperature of 0.019 dS m1 °C1 (Rhoades et al., 1999). Similar slopes have been measured by others (e.g., Persson and Berndtsson, 1998) for clay soils. Also, there is no reason to think that bulk electrical conductivity in saturated clayey soils is wholly determined by the electrolytic conductivity of the bulk soil water.
|
a and Effective Frequency as Covariates
|
|
a in the calibration model (Eq. [6]) resulted in some improvement (decrease) in the RMSE of regression (Table 7) compared with that for the model including only travel time (Table 3). The regression coefficient c was significant (P < 0.0001) and negative, in accordance with the theory (embodied in Eq. [5]) that increases in
a result in corresponding increases in apparent permittivity that are not related to increases in water content. All coefficients were significant (P < 0.0001) for individual soil calibrations as well, but contrary to theory, the c coefficient was positive for Soil B, and it was twice as large as that for Soil A. Also, the b coefficient for this soil was much smaller than those found by Topp and Reynolds (1998), by Ledieu et al. (1986), and from our own data (Table 3).
|
a and the effective frequency, fvi, in the calibration model (Eq. [7]) resulted in greater improvement in the RMSE of regression for the combined data, reducing it by one-half, even though no additional coefficient was fitted. Although RMSE values for Soils B and C were not reduced over those obtained with the Eq. [6] model, the c coefficients were all negative, in agreement with theory. Also, the b coefficients were similar in value for the combined data and for the individual soils, indicating that the Eq. [7] model encompassed the important physical effects of bulk electrical conductivity and signal frequency loss for the three soils and the TDR systems with varying cable lengths.
Surprisingly, inclusion of
a effects in Eq. [6] did not result in an overall decrease in the temperature dependency of estimated water contents (Table 8). Compared with estimates of water content from the Topp et al. (1980) equation, the calibrations based on Eq. [6] resulted in decreased temperature dependency for Soil A, but increased dependency for Soils B and C. Because of the large temperature dependency of
a, this was not expected. However, inclusion of both
a and effective frequency in the Eq. [7] model resulted in calibrations that exhibited uniformly small temperature dependencies, all <0.0006 m3 m3 °C1 temperature change. We conclude that the full model (Eq. [7]) is more physically realistic.
|
| DISCUSSION AND CONCLUSIONS |
|---|
|
|
|---|
The success of the Eq. [7] calibration model contradicts the analysis of Wraith and Or (1999) that argued for a negligible effect of the loss tangent, which includes
a/(2
fvi
o), for soil water with a conductivity of 0.075 S m1. That analysis used a conductivity that was too small to reflect conditions of our study. Soil BEC may be a factor of from 5 to 18 times smaller than soil solution EC (Rhoades et al., 1999). If the factor is 10 for our soils, then our BEC values ranging up to 0.2 S m1 would reflect an equivalent soil solution EC of approximately 2 S m1, much larger than the value supposed by Wraith and Or (1999). Moreover, it is important to note that our soil was wetted with water of low conductivity; the soil BEC is largely a result of clay content, clay type, and water content, but not the conductivity of the bulk soil solution per se. Analysis based on the loss tangent of soil water would not necessarily apply. We also note that the effective frequencies we measured for saturated soils were on the order of 0.1 GHz, well into the dispersive range for clay soils with large surface area and ion exchange capacity, such as bentonite (Robinson et al., 2003).
The full model (Eq. [7]) still does not include relaxation effects on the imaginary permittivity,
''. Relaxation effects may explain the slightly smaller values of coefficients b and c for Soil B, which was the most lossy soil. However, the success of the full model using combined data from three soils indicates that relaxation effects are minor in these soils.
The inclusion of effective frequency in the calibration model allowed both the low-pass filtering effect of longer cables and the decrease of effective frequency with temperature increase to be accounted for, practically eliminating both the tendency of the TDR system to overestimate water contents from probes attached to longer cables and the temperature dependency of TDR readings.
The Eq. [7] calibration model, including bulk electrical conductivity and effective frequency properties as independent variables, can be applied easily using data collected by the TACQ TDR data acquisition program. Modification of other TDR data acquisition systems to output the required slope data, which is already internally computed in these systems, should be easy. Most TDR systems already provide the needed data for BEC calculations. An important result of this study is that soil temperature need not be measured. Its effects are embedded in the behavior of BEC and effective frequency. The Eq. [7] model should be tested on a wider variety of soils; temperature, soil solution EC, and wetness ranges; and cable lengths and types. The fact that it does not explicitly include relaxation effects may not be important, as these may be inherent in the effective frequency reduction.
More problematic for wide adoption is the proposed calibration model's inability to predict a decrease in permittivity with temperature, a phenomenon reported by Wraith and Or (1999) and others for some soils. If the decrease is due to the decline of the permittivity of bulk water as temperature increases, then it may be that a summed effect will suffice to extend the proposed model to these soils. Such an effect would easily explain the decline in bulk soil permittivity reported by Wraith and Or (1999) for the 0 to 65°C temperature range and for the water contents in their study. Because we did not study a soil that behaved in this manner, study of the problem is beyond the scope of our investigation.
The fact that the proposed calibration model (Eq. [7]) does not reflect a complete physical analysis of the soil water system (e.g., does not explicitly account for relaxation losses, temperature effects on permittivity of bulk water) may cause it to be inappropriate for some soils. However, we believe that for many soils, it includes the important effects of frequency loss (whether due to cable length or soil dielectric) and bulk electrical conductivity. It will be interesting to see if calibrations in other soils using this model result in similar model parameters.
| ACKNOWLEDGMENTS |
|---|
| REFERENCES |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
S. R. Evett, R. C. Schwartz, J. A. Tolk, and T. A. Howell Soil Profile Water Content Determination: Spatiotemporal Variability of Electromagnetic and Neutron Probe Sensors in Access Tubes Vadose Zone J., November 17, 2009; 8(4): 926 - 941. [Abstract] [Full Text] [PDF] |
||||
![]() |
C.-C. Chung and C.-P. Lin Apparent Dielectric Constant and Effective Frequency of TDR Measurements: Influencing Factors and Comparison Vadose Zone J., July 7, 2009; 8(3): 548 - 556. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Moret-Fernandez, F. Lera, J. L. Arrue, and M. V. Lopez Measurement of Soil Bulk Electrical Conductivity Using Partially Coated TDR Probes Vadose Zone J., July 7, 2009; 8(3): 594 - 600. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. C. Schwartz, S. R. Evett, M. G. Pelletier, and J. M. Bell Complex Permittivity Model for Time Domain Reflectometry Soil Water Content Sensing: I. Theory Soil Sci. Soc. Am. J., May 1, 2009; 73(3): 886 - 897. [Abstract] [Full Text] [PDF] |
||||
![]() |
R.C. Schwartz, S.R. Evett, and J.M. Bell Complex Permittivity Model for Time Domain Reflectometry Soil Water Content Sensing: II. Calibration Soil Sci. Soc. Am. J., May 1, 2009; 73(3): 898 - 909. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Th. Mazahrih, N. Katbeh-Bader, S. R. Evett, J. E. Ayars, and T. J. Trout Field Calibration Accuracy and Utility of Four Down-Hole Water Content Sensors Vadose Zone J., August 1, 2008; 7(3): 992 - 1000. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. A. Huisman, C. P. Lin, L. Weihermuller, and H. Vereecken Accuracy of Bulk Electrical Conductivity Measurements with Time Domain Reflectometry Vadose Zone J., April 14, 2008; 7(2): 426 - 433. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. S. Seyfried and L. E. Grant Temperature Effects on Soil Dielectric Properties Measured at 50 MHz Vadose Zone J., October 8, 2007; 6(4): 759 - 765. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. M. Regalado, A. Ritter, and R. M. Rodriguez-Gonzalez Performance of the Commercial WET Capacitance Sensor as Compared with Time Domain Reflectometry in Volcanic Soils Vadose Zone J., April 9, 2007; 6(2): 244 - 254. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. A. Huisman and H. Vereecken Comments on "Time Domain Reflectometry Laboratory Calibration in Travel Time, Bulk Electrical Conductivity, and Effective Frequency" Vadose Zone J., October 3, 2006; 5(4): 1071 - 1072. [Full Text] [PDF] |
||||
![]() |
S. R. Evett, J.A. Tolk, and T.A. Howell Response to "Comments on 'TDR Laboratory Calibration in Travel Time, Bulk Electrical Conductivity, and Effective Frequency'" Vadose Zone J., October 3, 2006; 5(4): 1073 - 1075. [Full Text] [PDF] |
||||
![]() |
S. R. Evett, J. A. Tolk, and T. A. Howell Soil Profile Water Content Determination: Sensor Accuracy, Axial Response, Calibration, Temperature Dependence, and Precision Vadose Zone J., July 26, 2006; 5(3): 894 - 907. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. R. Evett and G. W. Parkin Advances in Soil Water Content Sensing: The Continuing Maturation of Technology and Theory Vadose Zone J., November 11, 2005; 4(4): 986 - 991. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| The SCI Journals | Agronomy Journal | Crop Science | |||
| Journal of Natural Resources and Life Sciences Education |
Soil Science Society of America Journal | ||||
| Journal of Plant Registrations | Journal of Environmental Quality |
The Plant Genome | |||